Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
67
5 Place the code key reader on a level surface such
as a bench. Hold the meter 4-6 inches above the
code key reader so that a connection can be made
between the two infrared windows.
6 Touch
to begin downloading data.
n The code le is ready for transmission as long as
the LED on the code key reader is ashing green,
even if the code key is removed.
After the LED stops ashing, remove the code
key and insert a new code key for download, if
required.
If you get an error message that the download
was not successful, re-insert the same code key
and try again.
n Do not exchange code keys while the code key
reader is still ashing. If the code key reader is still
ashing, it will continue to transmit the previously
loaded code le and ignore the code le on the
newly inserted code key. You may get an error
message on the meter.
Once the connection is made, the meter provides you
with status information on the download.
Vedere la pagina 66