General Product Information
150
Subject/Attribute Range Default Device DMS
QC time of day 6 POCT1-A data elds
00:00 – 23:59
06:00
09:00
12:00
15:00
18:00
22:00
N Y
QC time of day set 6 POCT1-A data elds
0 or 1
1 0 0 0 0 0 N Y
Result deletion
Result deletion algorithm. Automatic or
First in First out.
0: automatic
1: FIFO
1 N Y
Result retention time (number of days
since a result was measured before it can
be deleted automatically if result deletion
algorithm is set to “automatic” and no
more memory is available)
1 – 1,000 d 30 N Y
Result download required 0: disabled
1: enabled
0 N Y
Electronic communication
Download warning 0 – 999 h 0: disabled N Y
Download Lockout 0 – 999 h 0: disabled N Y
Maximum number of list items transferred
in one POCT1-A message
1 – 500 75 N Y
Application timeout (within this time the
application expects a response by the
DMS to any POCT1-A command)
5 – 6,000 s 60 N Y
WLAN settings (general)
WLAN enabled 0: disabled
1: enabled
1 N Y
Use DHCP 0: disabled
1: enabled
0 N Y
IP (the meter’s static IP-address if no DHCP
is used)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Subnet Mask (subnet mask to be used by
the meter if static IP is used. Note: if DHCP
is used; the DHCP-subnet mask will be
used.)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Use DNS 0: disabled
1: enabled
0 N Y
DMS host (the host’s DNS name if DNS is
used)
0 – 20 characters “” N Y
DMS IP address (the host’s IP address if no
DNS is used)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y