General Product Information
146
Subject/Attribute Range Default Device DMS
Comments requirement level: if out of ...
(only valid if Comments required = 1)
0: normal range
1: critical range
2: reportable range
3: measurement range
0 N Y
Custom comments 0: enabled
1: disabled
0 N Y
Control lot verication 0: display only
1: yes / no conrmation
2: prompt for entry
3: scan only
1 Y Y
Control result display 0: value (numeric)
1: PASS / FAIL
0 N Y
Critical range display 0: value (numeric)
1: HI / LO
0 N Y
Critical range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y Y
Critical range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y Y
Critical range message enabled Whether to display the
out-of-critical range
warning message (1) or
not (0)
1 N Y
Critical range text 0 – 100 characters “” N Y
Normal range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y Y
Normal range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y Y
STAT tests allowed 0: no
1: yes
1 Y Y
Number of STAT tests allowed 0 – 9 9 Y Y
Reportable range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y Y
Reportable range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y Y
Reportable range message Whether to display a
reportable range
message (1) or not (0)
1 N Y
Reportable range message text 0 – 100 characters “” N Y
Strip limitations warning Conguration, whether to
display the strip
limitations warning (1) or
not (0).
0 Y Y