General Product Information
151
Subject/Attribute Range Default Device DMS
DMS port (the port number to request
open socket from the DMS. Roche
recommends use of port numbers ≥ 1024.)
0 – 65,535 0 N Y
DNS IP (IP address of the DNS-server if
DHCP is not used and DNS is congured)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Gate IP (static gateway IP address) 0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
WLAN settings (security)
Cipher Type (encyrption method) 0: None (no cipher type)
1: AES (symmetric key
cryptography)
2: TKIP (using dynamic
keys)
3: AES | TKIP (AES and
TKIP together)
4: WEP40 (WEP with 64
bit key length)
5: WEP104 (WEP with 128
bit key length)
0: disabled N Y
Security Type * 0: open (no security/
encryption)
1: WEP
2: WPA_PSK (WPA with
pre shared key)
3: WEP2
4: EAP**
0 N Y
* Note: Only those combinations of cipher type and security type are allowed that are listed below, see table
“Allowed combinations of cipher and security type”.
**When using option 4 ensure that an appropriate WLAN EAP package has already been downloaded from
the DMS to the meter. Within this EAP container, the appropriate EAP type (TLS, PEAP or TTLS) and the
diverse EAP conguration settings have to be congured).
SSID 0 – 32 characters “” N Y
WEP-key (40 or 104bits) NUL-terminated string of
10 or 26chs (HEX)
„00000
00000“
N Y
WEP-authentication 0: open system
authentication
1: shared key
authentication
0 N Y
WPA Key Type 0: Passphrase
1: Key
0 N Y
Commenti su questo manuale