Introduction
20
To avoid ESD, do not use the meter in a very dry
environment, particularly one in which synthetic materials
(e.g., carpets) are present that might cause damaging
static discharges, or near electronic equipment that is a
source of electromagnetic radiation.
Local Area Network: protection
from unauthorized access
If this product is connected to a local area network, this
network must be protected against unauthorized access.
In particular, it must not be linked directly to any other
network or the Internet. Customers are responsible for the
security of their local area network, especially in
protecting it against malicious software and attacks. This
protection might include meausres, such as a rewall, to
separate the device from uncontrolled networks as well
as measures that ensure that the connected network is
free of malicious code.
Wired network connection
If connected to a local area network, the ACCU-CHEK
Inform II Base Unit or ACCU-CHEK Inform II Base Unit
HUB must be protected against unauthorized access. For
additional information about the ACCU-CHEK Inform II
Base Unit or ACCU-CHEK Inform II Base Unit HUB, refer
to the ACCU-CHEK Inform II Base Unit HUB Operator’s
Manual.
Wireless network connection
(RF card)
The optional RF card for a wireless network connection
allows the meter to send data (test results, patient IDs,
operator IDs, etc.) to the data management system
with out the need to return the meter to the base unit.
This feature must be congured by the system
administrator. Observe the guidelines of your facility for
using wireless local area network connections. For
information about how to temporarily enable or disable
this function, see page 33. For a description of the
ACCU-CHEK Inform II meter’s ability to connect to
Wireless Local Area Net works (WLAN, Wi-Fi), see
appendix B.