Patient Glucose Testing
44
The following messages may appear in addition to the
standard test result:
CR LO (below the Critical Range threshold, but
within the Reportable Range)
CR HI (above the Critical Range threshold, but
within the Reportable Range)
RR LO (below the Reportable Range threshold, but
within the System Measurement Range)
RR HI (above the Reportable Range threshold, but
within the System Measurement Range)
LO (below the System Measurement Range)
HI (above the System Measurement Range)
Instructions on how to add comments are provided in the
following section.
If you do not wish to add a comment to the test result,
touch to return to the Main Menu.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives for disposal of
potentially infectious samples and materials.
For test results that lie outside the critical or
reportable range, a message (up to 100 characters in
length) can be configured during setup. This message
is then displayed with the corresponding test results.
Test results are also saved when the meter is powered
off in the results screen or when it powers off
automatically.