General Product Information
137
12 General Product Information
Technical data
* Accu-Chek Inform II Base Unit Light: only red, green.
Specification Meter Base unit Power supply unit
Height 47 mm / 1.85 in 105 mm / 4.13 in 34 mm / 1.34 in +
AC plug (28-40 mm /
1.1-1.6 in)
Width 92 mm / 3.62 in 150 mm / 5.91 in 52 mm / 2.05 in
Length 190 mm / 7.48 in 150 mm / 5.91 in 88 mm / 3.46 in
Weight 350 g (with rechargeable
battery)
366 g (with rechargeable
battery AND RF card)
Accu-Chek Inform II
Base Unit
615 g with wall mount
Accu-Chek Inform II
Base Unit Light
573 g with wall mount
N/A
User interface Touchscreen and
barcode scanner
LED (tricolor: red, green,
blue) *
N/A
Display resolution
(touchscreen)
320 x 240 pixel
single color
N/A N/A
Memory 2,000 Results
5,000 Operator IDs
4,000 Patient IDs
N/A N/A
Operating temperature 3 to 50 °C
37 to 122 °F
3 to 50 °C
37 to 122 °F
0 to 40 °C
32 to 104 °F
Measurement
temperature
Test strip dependent:
Accu-Chek Performa test strips: 6 to 44 °C /43 to 111 °F
For Accu-Chek Performa test strips with the green symbol and for Accu-Chek
Inform II strips, refer to the respective package insert.
Storage temperature
(long-term storage)
2 to 25 °C
36 to 77 °F
-25 to 69 °C
-13 to 156 °F
-40 to 70°C
-40 to 158 °F
Humidity (operating) 90% RH at 32 °C / 90 °F
(non-condensing)
N/A N/A
Air pressure 0.7 to 1.06 bar
70 to 106 kPa
0.7 to 1.06 bar
70 to 106 kPa
N/A
Relative humidity
(storage)
< 95% < 95% N/A