Pagina 1 - Accu-Chek Inform II Training

5 April 2012 page 1 © 2009 Roche5 April 2012 page 1 © 2009 RocheAccu-Chek Inform II TrainingKeryn Smith, Roche April 2012Bernice Smith and Anne-Marie

Pagina 2

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemQuality Control• Control 1 and 2 solutions run daily • For

Pagina 3

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemCalibration • The process of testing and adjusting values o

Pagina 4 - Cobas IT 1000 Software

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemCalibration VerificationKeep strips in the original vial Al

Pagina 5 - Patient

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemTraining and certification• Operator ID information is stor

Pagina 6 - Connectivity

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemTraining and certification• Only trained operators will hav

Pagina 7 - Quality Assurance

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemTraining and certificationInformation is imported into coba

Pagina 8 - Aim of the policy

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemDuring this period we are using a test environment. Please

Pagina 9 - Related documents

Blood Glucose Monitoring using Point of Care testing at HVDHB

Pagina 10 - Lockout if not done

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemTraining and certification • Training period is for 3 week

Pagina 11 - Calibration

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemTraining Period till 30thApril• Trainer (you) trains staff

Pagina 12 - Calibration Verification

Bernice Smith, POCC, HVDHB HVDHBJennie Johnson, HVDHB Programme Manager, Information ServicesHVDHBStewart Clark, Lab Manager, HVDHB HVDHBWillie Andrew

Pagina 15

Your role asAccu-Chek Inform II trainer • Train staff in the use of the Accu-Chek Inform II in accordance to the policyAccu-Chek Inform II operator •

Pagina 16

HELPThe Laboratory keeps back-up meters 24/7. The same swap out procedure will be in place.• Bernice Smith• POCT Co-Ordinator• Ext 8660 Wed and Fri o

Pagina 17

5 April 2012 page 24 © 2009 RocheThe Accu-Chek Inform II SystemOverview• First wireless hospital glucose meter• Robust design, seamless casing able to

Pagina 18

5 April 2012 page 25 © 2009 RocheThe Accu-Chek Inform II SystemOverview1. Test strip port Insert the t

Pagina 19

5 April 2012 page 26 © 2009 RocheThe Accu-Chek Inform II SystemOverview4. Barcode scanner (laser) The integrat

Pagina 20

5 April 2012 page 27 © 2009 RocheThe Accu-Chek Inform II Base Unit/Base Unit LightOverviewBoth versions of the base unit have the following elements:1

Pagina 21

5 April 2012 page 28 © 2009 RocheThe Accu-Chek Inform II SystemData Transfer - Wireless• Results transferred as soon as a measurement has been complet

Pagina 22 - Your role as

5 April 2012 page 29 © 2009 RocheThe Accu-Chek Inform II SystemData Transfer - Wired• Data communicated as soon as the meter is docked• Every 10 minut

Pagina 23

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II System

Pagina 24 - Overview

5 April 2012 page 30 © 2009 RocheThe Accu-Chek Inform II Test StripOverview• Same test strip used by 80% of people with Diabetes in the community (Per

Pagina 25

5 April 2012 page 31 © 2009 RocheThe Accu-Chek Inform II Test Strip Limitations• If Peripheral circulation is impaired, collection of capillary blood

Pagina 26

5 April 2012 page 32 © 2009 RocheThe Accu-Chek Inform II SystemPowering Up• BloodIn the Power Up screen you change the Contrast, activate/deactivate R

Pagina 27

Quality Control Testing with the Accu-Chek Inform IIWhy test the Performa Quality Control Solution?To be confident that I know my Accu-Chek Inform II

Pagina 28 - Data Transfer - Wireless

Quality Control Testing with the Accu-Chek Inform IIPerforma Quality Control Solution• Has defined (known) values Level 1 (L1): Lo (low values in tes

Pagina 29 - Data Transfer - Wired

Quality Control Testing with the Accu-Chek Inform IIQC Lockout• QC Lockout is activated after 24 hours• A QC result must be performed successfully be

Pagina 30

Testing with the Accu-Chek Inform IIPerforming a Quality Control Test1. Power up the meter by pressing the power button2. Touch the or wait 5 s

Pagina 31 - Limitations

Testing with the Accu-Chek Inform IIPerforming a Quality Control Test5. Scan in the barcode on the Control vial for desired level 6. Scan the strip lo

Pagina 32 - Powering Up

Testing with the Accu-Chek Inform IIPerforming a Quality Control Test10. Control solution is pulled into the test strip by capillary action. The mete

Pagina 33

Testing with the Accu-Chek Inform IIQuality Control TestingShould the result be outside the specified target range, proceed as follows: • Suspend all

Pagina 34

Cobas IT 1000 SoftwarePatient Details for positive patient IDBi directional transferQC results(control 1 and 2)Patient ID and results, time, date of t

Pagina 35 - QC Lockout

5 April 2012 page 40 © 2009 RocheTesting with the Accu-Chek Inform IIPreparing to Test1. Operator to wash hands.2. Identify patient.3. Select, prepare

Pagina 36

5 April 2012 page 41 © 2009 RocheTesting with the Accu-Chek Inform IIPreparing to Test7. Hold finger firmly 8. Position the lancet on the site and d

Pagina 37 - AFTER REMOVING A STRIP

Testing with the Accu-Chek Inform IIPerforming a Blood Glucose Test1. Power up the meter by pressing the power button2. Touch the or wait 5 sec

Pagina 38

Testing with the Accu-Chek Inform IIPerforming a Blood Glucose Test5. Enter the patients NHI Number using the keypad and touch the key. Verif

Pagina 39 - Quality Control Testing

Testing with the Accu-Chek Inform IIPerforming a Blood Glucose Test10. Blood is pulled into the test strip by capillary action. The meter beeps and a

Pagina 40 - Preparing to Test

The Accu-Chek Inform II SystemMemoryMeter stores 2000 patient, test results and maintenance logs results alongside:• Patient ID, glucose control, or s

Pagina 41

The Accu-Chek Inform II SystemAccessing the Memory1. Touch Review Results2. Results are in sequential order and grouped by day. Touch or

Pagina 42

The Accu-Chek Inform II SystemMaintenance and CareStorage and Handling• Handle the meter and its system components carefully. Avoid dropping it or ban

Pagina 43

The Accu-Chek Inform II SystemMaintenance and CareCleaning and Disinfection• Use only HVDHB approved cleaning product only to clean and disinfect the

Pagina 44

Infection prevention and control • 10.4 Procedure for use and cleaning of equipment used on patients in Transmission-Based Precaution isolation rooms•

Pagina 45

Blood Glucose Monitoring using Point of Care testing at HVDHBPatient results

Pagina 46 - Accessing the Memory

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemDamaged meters Under warranty unless Abused; where the war

Pagina 47 - Maintenance and Care

Key points to remember with patterns of monitoring• A recent HBA1c is useful• Blood glucose testing frequency should be charted• Appropriate monitorin

Pagina 48

Recommended NICE guidelines for self –monitoring• Type 1 diabetes The majority of people with Type 1 diabetes in hospital need to monitor their blood

Pagina 49

intensive monitoring and maintenance monitoringpatterns•The usual aim is to obtain a profile of glucose levels across the day. The number of tests wil

Pagina 50 - Damaged meters

Factors altering target range for individual patients• Child under 6 years• Elderly • Illness• Infection• Stress• Psychological problems• Occupationa

Pagina 51 - • A recent HBA1c is useful

Critical Action Points• The ‘critical’ action points that will be displayed on the Accu-Chek Inform II meter are set for below 4.0mmol/L and above 20m

Pagina 52

Key points to remember with patterns of monitoring• A recent HBA1c is useful• Blood glucose testing frequency should be charted• Appropriate monitorin

Pagina 53

Patients who manage their own blood glucose monitoringEg Patient attending Outpatient’s clinic, physiotherapy ectIs the exclusion – to this Policy

Pagina 54

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II System• Trainer (you) trains staff in ward. • Operators then use

Pagina 55 - Critical Action Points

5 April 2012 page 59 © 2009 RocheThank you for your attention.Roche Diagnostics Ltd.6343 RotkreuzSwitzerlandCOBAS and LIFE NEEDS ANSWERS are trademark

Pagina 56

Blood Glucose Monitoring using Point of Care testing at HVDHBConnectivity

Pagina 57

Quality Assurancea) POCT has the same level of quality assurance as is provided for all testing performed within theHVDHB Laboratory.b) All POCT compl

Pagina 58

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II System• To ensure quality and reliability of glucose results meas

Pagina 59 - Thank you for your attention

Blood Glucose Monitoring using Point of Care testing at HVDHBThe Accu-Chek Inform II SystemPoint of Care TestingManagement of High Blood Glucose for p