Roche ACCU-CHEK Inform II Manuale dell'Operatore

Navigare online o scaricare Manuale dell'Operatore per Misurazione, test & controllo Roche ACCU-CHEK Inform II. Roche ACCU-CHEK Inform II Operator`s manual Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa

Sommario

Pagina 1 - ACCU-CHEK

Inform IIACCU-CHEK ®BLOOD GLUCOSE MONITORING SYSTEMOperator’s Manual

Pagina 2

Introduction10If you need help Information about using the system, the screen menus and performing a test can be found in this operator's manual.

Pagina 3 - Inform II System

Initial Startup100This display is visible when no communication takes place. This display is visible when software updates are trans-ferred to the met

Pagina 4 - 7.5V 1.7A

Initial Startup101Initial setup on the meterThere are two ways to customize the setup of the meter: directly on the meter (see the following menu over

Pagina 5

Initial Startup102Menu overview The following tables contain a brief overview of the menu structure. The menus can be used to operate the meter fully

Pagina 6

Initial Startup103Main Menu 2Maintenance – Add Comment– Maintenance ResultProficiency – Enter Sample ID– Verify Strip Lot– Insert Test Strip– Apply Sa

Pagina 7

Initial Startup104Admin MenuAdmin – Language– Date/Time– SetupLanguage – German– French– Spanish– Italian– Dutch– Swedish– English – Danish– Portugues

Pagina 8

Initial Startup105Opening the Setup MenuAll settings described here are configured via the Setup Menu. To open the Setup Menu, proceed as follows:1 To

Pagina 9 - Before you start

Initial Startup106Date and time formatUse this menu to select the date and time format for the display. You can also choose whether the operator is al

Pagina 10

Initial Startup107Display options and optional testsUse this menu to select the unit of measure for test results and enable or disable optional tests.

Pagina 11

Initial Startup108Options for test stripsThis menu allows you to select options for handling lot numbers and specify whether the operator is allowed t

Pagina 12

Initial Startup109On the second settings screen, you can select options for handling lot numbers with glucose control and linearity tests. 1 Touch the

Pagina 13

Introduction11What can the system do for you? The Accu-Chek Inform II system has the following fea-tures and properties: Perform patient blood glucos

Pagina 14

Initial Startup110Options for glucose control testsThis menu contains options that allow you to specify whether and at what intervals glucose control

Pagina 15

Initial Startup111Value ranges (normal, critical, reportable)This menu allows you to set limit values for results as Normal, Critical, or Reportable.

Pagina 16

Initial Startup112Options for Operator ID EntryThis menu allows you to specify if and how operator login is performed. 1 Touch the desired option to e

Pagina 17

Initial Startup113Patient ID optionsThis menu allows you to specify the criteria for entering a patient ID. 1 Touch the desired option to enable it:

Pagina 18

Initial Startup114Creating a setup passwordThis screen allows you to create a password for all the settings described in this chapter. The password en

Pagina 19

Initial Startup115Setting the date and timeThis setting can be hidden or require entry of a password, based on configuration.1 Touch in the Main Men

Pagina 20

Initial Startup116Beeper optionsThis setting can be used to set the volume of the beeper.1 Touch in the Main Menu to open the Main Menu 2 screen. 2

Pagina 21 - System components

Initial Startup117Diagnostics viewUnder Diagnostics you can find information about the system, such as software version, number of data records stored

Pagina 22 - Overview of the meter

Initial Startup118Unlocking a Download LockoutWhen using a data management system for configura-tion, a Download Lockout can be configured. This locko

Pagina 23

Maintenance and Care11910 Maintenance and CareConditions for storage and shippingGeneral operating conditions Please observe the following points to e

Pagina 24 - Overview of the base unit

Introduction12Safety information and additional informationThis section explains how safety-related messages and information related to the proper han

Pagina 25 - Reagents and consumables

Maintenance and Care120Storage Store the system and test strips in the same environment in which they are used. Do not store the meter in direct su

Pagina 26

Maintenance and Care121Cleaning the meter Wipe the surfaces with a soft cloth slightly dampened (not wet) or you may spray the meter per the directio

Pagina 27 - Powering up the meter

Maintenance and Care122Cleaning the base unit Unplug the base unit before cleaning. Wipe the surfaces with a soft cloth slightly dampened (not wet)

Pagina 28 - Low Power Mode after:

Maintenance and Care123Cleaning the accessory box Wipe the surface with a soft cloth slightly dampened (not wet). If spraying the accessory box, pla

Pagina 29

Maintenance and Care124DisinfectingHealthcare professionals must wear gloves and follow their institution’s infection control procedures when han-dlin

Pagina 30 - Entering the operator ID

Maintenance and Care125Disinfecting the meter Remove the meter from the base unit prior to disinfecting. Lay the meter on a flat surface while wipi

Pagina 31

Maintenance and Care126WARNINGDo not spray directly into the strip port! Moisture in the strip port may lead to incorrect blood glucose results. If yo

Pagina 32

Maintenance and Care127Disinfecting the base unit Remove the meter from the base unit prior to disinfecting. Unplug the base unit before disinfecti

Pagina 33 - 3 Patient Glucose Testing

Maintenance and Care128Disinfecting the accessory box Be certain that solutions contain acceptable active ingredients for disinfecting the system. If

Pagina 34 - ■ Perform the test

Maintenance and Care129Logging maintenance activitiesCleaning and other maintenance activities can be logged in the meter. Make sure that all cleaning

Pagina 35

Introduction13Important information regarding safetyDisposal of the systemOperator qualification Only trained healthcare professionals may operate the

Pagina 36

Maintenance and Care130

Pagina 37

Troubleshooting13111 TroubleshootingThe Accu-Chek Inform II meter continually checks its sys-tems for unexpected and unwanted conditions. A troublesho

Pagina 38

Troubleshooting132Test result LO/HI The glucose result may be below (LO) or above (HI) the measurement range of the system.– Refer to the package inse

Pagina 39

Troubleshooting133Base unitLED is not illuminated Not connected to power supply unit or power supply unit is defective, base unit is defective, or pow

Pagina 40 - Strip Lot 545794

Troubleshooting134Error messages All error messages displayed by the system are accompa-nied by a description of the error and a possible solution. Ta

Pagina 41

Troubleshooting135– E: Error; touch to confirm. To resolve the problem, take the actions suggested.– W: Warning; touch to confirm.Sample message t

Pagina 42

Troubleshooting136Meter reset A meter reset can be performed to correct a number of unspecified errors (e.g., “frozen” screen, etc.). Use the fol-lowi

Pagina 43

General Product Information13712 General Product Information Technical data * Accu-Chek Inform II Base Unit Light: only red, green.Specification Meter

Pagina 44

General Product Information138* Not applicable to Accu-Chek Inform II Base Unit Light Specification Meter Base unit Power supply unitBattery voltage/t

Pagina 45

General Product Information139Further InformationOrdering Accu-Chek Inform II Operator’s Manual and Quick Reference GuideFor additional copies of the

Pagina 46 - Patient Glucose Testing

Introduction14Product safety Observe the following information to ensure product safety: The system is suitable for continuous operation.General car

Pagina 47 - 4 Glucose Control Testing

General Product Information140Information about software licensesThis product incorporates software modules developed under open source licenses. The

Pagina 48 - QC Due: Immediately

Appendix141A AppendixTable of configuration optionsThis section provides an overview of all the settings avail-able. The two rightmost columns describ

Pagina 49

Appendix142Critical range HI limit 10 – 600 mg/dL0.6 – 33.3 mmol/L600 mg/dL33.3 mmol/LYYCritical range LO limit 10 – 600 mg/dL0.6 – 33.3 mmol/L10mg/dL

Pagina 50

Appendix143Formats and LanguageDate format 1: MM/DD/YY2: DD.MM.YY1YYTime format 1: 24 hours 2: 12 hours2YYLanguage setting * 1: German3: French4: Span

Pagina 51

Appendix144Operator IDOperator ID entry control 0: None1: Prompt2: Scan only3: Prompt (numeric only)1YYOperator ID entry control on Glucose control on

Pagina 52

Appendix145Patient ID validation mode 0: none1: length2: list3: list allowing entry if not on list4: length if numeric0YYPatient ID length(Allowed cha

Pagina 53

Appendix146Result deletionResult deletion algorithm 0: automatic1: FIFO1NYResult retention time (number of days since a result was measured before it

Pagina 54 - Control L2 (Hi) 123456

Appendix147WLAN settings (security)Cipher Type (encyrption method) 0: None (no cipher type)1: AES (symmetric key cryptography)2: TKIP (using dynamic k

Pagina 55

Appendix148Subject/Attribute Range Default Device DMSObserved Test Sequence (OTS)OTS functionality * 0: disabled1: enabled0NY* Depends on available DM

Pagina 56

Appendix149Allowed combinations of cipher and security typeOperator and patient ID barcode maskssecurity_type cipher_typewep_auth_typewep_keywpa_key_t

Pagina 57

Introduction15Battery Pack The meter contains a rechargeable battery pack that begins charging as soon as it is placed in an active base unit (i.e., o

Pagina 58 - Glucose Control Testing

Appendix150Example of barcode symbologies– Codabar – Code 39– Code 93– Code 128– EAN 13– Interleaved 2 of 5

Pagina 59 - 5 Review Results

Appendix151B AppendixOption: Wireless network (WLAN)Application area The optionally available module for wireless network con-nectivity (RF card) allo

Pagina 60

Appendix152

Pagina 61

Index153IndexAAccessory boxCleaning ... 123Disinfecting ...

Pagina 62

Index154IIconsIdentification plate ...4Packaging ...

Pagina 63

Index155SSafetyProtection against infection ...13User qualification ...13Safety informa

Pagina 64

Index156

Pagina 65

Notes157Notes

Pagina 66

Notes158Notes

Pagina 68

Introduction16TouchscreenLaser scanner The built-in barcode scanner emits a laser beam when activated.The built-in barcode scanner is a Class 1 laser,

Pagina 69

ACCU-CHEK, ACCU-CHEK INFORM, ACCU-CHEK PERFORMA and COBAS are trademarks of Roche. Roche Diagnostics GmbHSandhofer Strasse 11668305 MannheimGermany

Pagina 70

Introduction17Radiofrequency radiation exposure informationNo operating limitations apply to meters without an RF card or to meters with an RF card th

Pagina 71

Introduction18This device complies with Part 15 of the FCC Rules and with RSS-210 of Industry Canada. Operation is subject to the following two condit

Pagina 72

Introduction19Note on systems with older hardwareThe Accu-Chek Inform II system complies with the emis-sion and immunity requirements described in IEC

Pagina 73

2Manual version Revision date ChangesVersion 1.0 2008-01 New documentVersion 2.0 2009-08 Update, new SW version 02.00Version 3.0 2010-09 Update, new S

Pagina 74

Introduction20This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. T

Pagina 75

Introduction21System componentsThe Accu-Chek Inform II system includes the following components and accessories: A MeterB Code key readerC Base unit w

Pagina 76

Introduction22Overview of the meterThe meter has the following elements: 1 Test strip portInsert the test strip here.2 Touchscreen(touch-sensitive dis

Pagina 77 - 7 Linearity Testing

Introduction23Overview of the code key readerTest strip vials include a code key.1 This code key is read by the code key reader and the data is sent t

Pagina 78

Introduction24Overview of the base unitTo provide flexibility in line with customer requirements, two versions of the base unit are available. The A

Pagina 79 - Performing a linearity test

Introduction25Electrical connections are located on the back of the base unit (for a better view the removable wall mount is shown as transparent). Fo

Pagina 80

Introduction26Instructions for initial setupThe meter must be configured prior to initial use. During this setup, the following parameters are configu

Pagina 81

Powering Up and Entering an Operator ID272 Powering Up and Entering an Operator IDPowering up the meter1 Press and release the On/Off button . The met

Pagina 82 - Linearity 12345678 L1

Powering Up and Entering an Operator ID28Adjusting the display Using the Display options, you can adjust display param-eters to your needs: Adjust th

Pagina 83

Powering Up and Entering an Operator ID29Enabling/disabling the RF card If the unit is equipped with an RF card, you can temporar-ily enable or disabl

Pagina 84 - Linearity Testing

©2010 Roche Diagnostics. All rights reserved. 0 4807839001 (03) 2010-09 ENOperator’s ManualAACCU-CCU-CHEKHEK ® Inform II System

Pagina 85 - 8 Proficiency Testing

Powering Up and Entering an Operator ID30Closing startup Once you have completed all the necessary changes, touch to proceed to the screen used to

Pagina 86

Powering Up and Entering an Operator ID31Entering an operator ID with barcode scannerWhen the screen for entering the operator ID is displayed:1 Press

Pagina 87 - Performing a proficiency test

Powering Up and Entering an Operator ID32Entering the operator ID manuallyWhen the screen for entering the operator ID is displayed:1 Touch the letter

Pagina 88

Patient Glucose Testing333 Patient Glucose Testing Information regarding blood glucose testingPreparing to test The following requirements must be met

Pagina 89 - Sample 123456789

Patient Glucose Testing34Performing a patient glucose testOverview of test procedure A patient glucose test comprises the following steps: Enter the

Pagina 90

Patient Glucose Testing35You now have three different options, depending on setup, for assigning the subsequent test to a patient.* Depending on the D

Pagina 91

Patient Glucose Testing36Entering the patient ID manually Use the displayed keypad to enter the patient ID. You can select characters in the same mann

Pagina 92 - Proficiency Testing

Patient Glucose Testing37Selecting the patient ID from a listChoose the patient ID from a list*, if a list has been down-loaded to the meter (from the

Pagina 93 - Connecting the base unit

Patient Glucose Testing38Entering a patient ID with barcode scannerWhen the screen for entering the patient ID is displayed:1 Press and release . The

Pagina 94

Patient Glucose Testing39Confirming or selecting the test strip lotOnce you have entered and confirmed the patient ID, you are asked to choose the lot

Pagina 95

4©2010, Roche Diagnostics. All rights reserved.The contents of this document, including all graphics, are the property of Roche Diagnostics. Informati

Pagina 96

Patient Glucose Testing40Inserting test strips After confirming the test strip lot, you are prompted to insert the test strip.1 Remove the test strip

Pagina 97

Patient Glucose Testing41Applying a blood sample Once the meter has detected the test strip, you are prompted to apply a blood sample. 1 Wait until th

Pagina 98

Patient Glucose Testing42Results screen The hourglass icon indicates the test is running. When the test is completed and the result is ready, the mete

Pagina 99 - Docking the meter

Patient Glucose Testing43 The System Measurement Range refers to the measurement range of the system itself (strips and meter) and is the only range

Pagina 100

Patient Glucose Testing44The following messages may appear in addition to the standard test result: CR LO (below the Critical Range threshold, but wi

Pagina 101 - Initial setup on the meter

Patient Glucose Testing45Adding comments You can add up to three comments to a test result. Com-ments can provide, for example, additional information

Pagina 102 - Initial Startup

Patient Glucose Testing46

Pagina 103

Glucose Control Testing474 Glucose Control TestingInformation regarding glucose control testsObserve the applicable regulations and directives of the

Pagina 104

Glucose Control Testing48Glucose control testing intervals Intervals between running glucose control tests are determined by your facility. These inte

Pagina 105 - Opening the Setup Menu

Glucose Control Testing49Information stored during glucose control testingThe following information is stored for every glucose con-trol test using co

Pagina 106 - Date and time format

51 Introduction 9Before you start...

Pagina 107 - ■ Enable Tests

Glucose Control Testing50Performing glucose control testsOverview of test procedure A glucose control test using control solution comprises the follow

Pagina 108 - Options for test strips

Glucose Control Testing51Starting a glucose control test After preparing the meter as described, you can proceed to the steps directly related to cont

Pagina 109 - ■ Control Lot Verification

Glucose Control Testing52Confirming or selecting the lot number for control solutionsOnce you have selected the level, you are asked to con-firm or en

Pagina 110 - ■ Control Frequency

Glucose Control Testing53Confirming or selecting the test strip lotOnce you have entered and confirmed the lot number of the control solution, you are

Pagina 111

Glucose Control Testing54Inserting test strips After confirming the test strip lot, you are prompted to insert the test strip.1 Remove the test strip

Pagina 112 - Options for Operator ID Entry

Glucose Control Testing55Applying the control solution Once the meter has detected the test strip, you are prompted to apply the control solution. 1 W

Pagina 113 - Patient ID options

Glucose Control Testing56Results screen The hourglass icon indicates the test is running. When the test is completed and the result is ready, the mete

Pagina 114 - Creating a setup password

Glucose Control Testing57Performing a STAT test The meter can be configured to allow a STAT patient glu-cose test to be run even if the meter is in QC

Pagina 116 - Beeper options

Review Results595 Review ResultsDisplaying test results from the memoryInformation stored in data records for test resultsWhen you retrieve the data r

Pagina 117 - Diagnostics view

63 Patient Glucose Testing 33Information regarding blood glucose testing ...

Pagina 118 - Unlocking a Download Lockout

Review Results603 Touch an entry in the list to display the related details.4 Touch Patient, if you wish to display results for a specific patient onl

Pagina 119 - 10 Maintenance and Care

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter616 Storing Test Strip, Control Solution, and Linearity Solution

Pagina 120 - Cleaning

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter62Transferring code key information to the meterThe following des

Pagina 121

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter635 Place the code key reader on a level surface such as a bench.

Pagina 122

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter64Information about the expiration date and parameters for contro

Pagina 123

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter65The parameters for control solutions consist of four sepa-rate

Pagina 124 - Disinfecting

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter66Once you have finished updating the test strip informa-tion, yo

Pagina 125

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter67Storing control solution informationGlucose control solution lo

Pagina 126

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter684 Select the level (L1/Lo or L2/Hi).5 Use the keypad to enter t

Pagina 127

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter69Once you have finished updating the control solution informatio

Pagina 128

77 Linearity Testing 77Information regarding linearity tests...

Pagina 129

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter70Selecting a stored lot number as the current lot numberYou can

Pagina 130 - Maintenance and Care

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter713 Touch the lot number you wish to select as the current lot nu

Pagina 131 - 11 Troubleshooting

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter72Storing linearity test informationObserve the applicable regula

Pagina 132 - Troubleshooting

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter734 Use the keypad to enter the lot number.5 Touch to confirm t

Pagina 133

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter74Selecting a stored lot number as the current lot numberYou can

Pagina 134 - To reject, touch

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter753 Touch the lot number you wish to select as the current lot nu

Pagina 135

Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter76

Pagina 136

Linearity Testing777 Linearity TestingInformation regarding linearity testsObserve the applicable regulations and directives of the responsible regula

Pagina 137 - Technical data

Linearity Testing78Information stored during linearity testingThe following information is stored for every linearity test: Test result Lot number o

Pagina 138 - General Product Information

Linearity Testing79Performing a linearity testOverview of test procedure A linearity test comprises the following steps: Check the lot number of the

Pagina 139 - Further Information

810 Maintenance and Care 119Conditions for storage and shipping ...

Pagina 140

Linearity Testing80Confirming or selecting the lot number for linearity test kitsYou are now prompted to confirm or enter the lot number of the linear

Pagina 141 - A Appendix

Linearity Testing81In the Linearity Test menu, the levels available for the lin-earity test are displayed. 5 Touch L1 to start the subsequent test wit

Pagina 142 - Appendix

Linearity Testing82Applying a linearity test sample Once the meter has detected the test strip, you will be prompted to apply the linearity solution.

Pagina 143

Linearity Testing83Results screen The hourglass icon indicates the test is running. When the test is completed and the result is ready, the meter beep

Pagina 144

Linearity Testing84

Pagina 145

Proficiency Testing858 Proficiency TestingInformation regarding proficiency testsObserve the applicable regulations and directives of the responsible

Pagina 146

Proficiency Testing86Information stored during proficiency testingThe following information is stored for every proficiency test: Test result Sample

Pagina 147

Proficiency Testing87Performing a proficiency testOverview of test procedure A proficiency test comprises the following steps: Enter a sample ID for

Pagina 148

Proficiency Testing88Entering the proficiency sample IDYou will now be asked to enter the sample ID. 1 Use the keypad to enter the sample ID, or2 Touc

Pagina 149

Proficiency Testing89Inserting test strips After confirming the test strip lot, you will be prompted to insert the test strip.1 Remove the test strip

Pagina 150

Introduction91 IntroductionBefore you startIntended use The Accu-Chek® Inform II system is intended for in vitro diagnostic use in the quantitative de

Pagina 151 - B Appendix

Proficiency Testing90Applying a proficiency sample Once the meter has detected the test strip, you are prompted to apply the proficiency sample. 1 Wai

Pagina 152

Proficiency Testing91Results screen The hourglass icon indicates the test is running. When the test is completed and the result is ready, the meter be

Pagina 153

Proficiency Testing92

Pagina 154

Initial Startup939 Initial StartupConnecting the base unit1 Slide the base unit upward and remove it from the wall mount (if in use). 2 Connect the po

Pagina 155

Initial Startup94Installing or replacing the battery packWhen shipped, the battery pack is not installed in the Accu-Chek Inform II meter. After insta

Pagina 156

Initial Startup95Removing the battery pack 1 If a battery pack is already installed, make sure that the meter is switched off.2 Place the meter face d

Pagina 157

Initial Startup96Installing the battery pack 1 Loosen the screws on the battery pack until they are protruding about 4-5 mm (2/10 in).2 Position the b

Pagina 158

Initial Startup975 Place the tapering end of the battery pack on the small ledge of the opening and push down like a lid to close (as shown in the ill

Pagina 159

Initial Startup988 Place a ruler across the back of the meter to check that the battery pack is properly positioned. The ruler should lie completely f

Pagina 160 - 0 4807839001 (03) 2010-09 EN

Initial Startup99Docking the meterDocking the meter in the base unit allows you to charge the battery. When docked, the meter shows different messages