Initial Startup
110
Options for glucose control tests
This menu contains options that allow you to specify
whether and at what intervals glucose control tests are
performed. If you require control tests, check your set-
tings for STAT tests (see page 107).
1 Touch the desired option to enable it:
■ Control Frequency
– Always OK: QC Lockout is disabled, patient tests
can be performed at any time. This applies
regardless of the result of running glucose control
tests.
– Last QC OK: QC Lockout is activated only when the
last glucose control test is out of range or a control
test is not available (e.g., when changing the test
strip lot).
– Hours: QC Lockout is activated after the specified
time. When a glucose control test is run
successfully, the counter is reset to “0”.
– Strip Count: QC Lockout is activated after reaching
the specified number of tests. When a glucose
control test is run successfully, the test strip
counter is reset to “0”.
– Shift: QC Lockout is activated per shift.
– Time Of Day: QC Lockout is activated at the
specified time of day. (Up to six times per day can
be configured.)
– Rotating QC: QC tests require alternating use of L1
and L2 controls, instead of requiring both controls
when performing a QC test.
2 Touch to save the settings. If the Hours or Strip
Count option is enabled, enter the desired number.
Or:
3 Touch to exit this menu without saving any
changes.
The display returns in both cases to the Setup Menu.
QC Frequency
03/16/10
12:48pm
Last QC OK
Always OK
Strip Count
Hours
Control Frequency:
Rotating QC
Shift
Time Of Day
123
456
789
0
QC Hours 12:48
pm
03/16/10
50
123
456
789
0
QC Strips 12:48
pm
03/16/10
50