Chapter 2: Control Testing 31 ACCU-CHEK Customer Care Service Center 1-800-858-8072
4. The time and date appear on the display with a
flashing test strip and blood drop symbol.
5. Remove the control bottle cap.
6. Apply one drop of the control solution to a flat,
clean, and easily washable surface such as the
palm of your hand, a plate, or a table. If you
are unsure if a surface is clean, wipe it off with
a slightly damp cloth or cotton swab. Hold the
meter so the test strip points downward. Gently
touch and hold the black notch on the end of
the test strip to the control solution. Do not put
control solution on top of the test strip. When
you see
displayed, you have enough control
solution in the test strip. Do not bend the test
strip while applying control solution.
7. Place the cap on the control bottle tightly.