Accu Chek Inform II Blood Glucose Meter Standard Operating
Procedure.
Authorised by: K. Ashton
Signature : ………….
Do not photocopy this document
Page 9 of 14
Last printed: 13 May 2015 ( ward-jones julie (rq6) rlbuht )
Reference Range/ Critical values
Any sample reading below 2.8 mmol/l or above 20 mmol/l, or which is not in keeping with the clinical
picture, should be verified by sending a venous specimen of blood to the laboratory for random blood
glucose testing.
If ‘LO’ is displayed on the meter, blood glucose may be below 0.6 mmol/L.
If ‘HI’ is displayed on the meter, blood glucose may be over 33.3 mmol/L.
A high or low result should be confirmed on a venous sample analysed in the laboratory.
If the blood glucose result does not reflect the patient’s clinical symptoms, or is unexpectedly High or Low, a
quality control test should be performed to check the performance of the meter. If this is satisfactorily, the
patient test should be repeated and if confirmed then medical staff informed.
Results Review
1. Press the Review Results button in the MAIN Menu.
2. Scroll the list using the arrow buttons on the right side of the screen.
3. Select specific results to view details (e.g. comments, flags, date and time).
4. View all results for a specific patient by pressing Patient in the main list and then entering a Patient
ID, or selecting a specific result for this patient in the main list and then pressing Patient.
5. Select only QC results by pressing the respective button.
Turnaround Time
All samples should be analysed immediately
Maintenance
Cleaning the Accu chek Inform II Meter and Base Unit
Personal protective equipment must be worn when performing any maintenance procedure.
The Inform11 Blood Glucose meter MUST be kept clean at all times. Any blood splashes MUST be
removed immediately using a soft cloth and a 1 in 10 dilution of bleach.
1. Remove the meter and unplug the base unit
2. Use a soft cloth slightly dampened (NOT WET) or pre moistened wipe with either soapy water or
1:10 dilution of bleach.
3. Carefully wipe the surfaces of the meter and base unit.
4. Dry the meter with a soft cloth or gauze.
Always thoroughly dry the meter and base unit before putting back into use or plugging in the
base unit.
5. Plug in the base unit and resume use of the system.
6. DO NOT immerse the meter or base unit in solution.
7. DO NOT spray the meter or base unit directly with solutions as this could cause the solution to enter
the case and damage the electronic components.
A soft cloth wetted with 70% isopropyl alcohol can be used to disinfect the surfaces of the meter. Ensure no
streaks remain on the touchscreen.
Pre moistened cleaning cloths (squeeze off excess cleaning solution) can be used to clean the surface of the
meter.
Ensure meter/base unit (including connectors) are thoroughly dried after cleaning or disinfecting.
Cleaning the scanner window
Use a clean, dry cloth to wipe the scanner window.