Performance Analysis
The performance data for the Accu-Chek Aviva Combo System (Accu-Chek Aviva Combo Meter with
Accu-Chek Aviva Test Strips) were obtained using capillary blood from diabetic patients (method comparison,
accuracy), venous blood (repeatability), and control solution (reproducibility). The system is calibrated with
venous blood containing various levels of glucose. The reference values are obtained using the hexokinase
method. For method comparison, the results were compared with results obtained using the hexokinase
method with deproteinization (automatic analyzer). The hexokinase method is traceable to an NIST standard.
Measuring Principle
Refer to your test strip package insert for more information.
w WARNING
• Strong electromagnetic fields may interfere with the proper operation of the meter. Do not use this
meter close to sources of strong electromagnetic radiation.
• To avoid electrostatic discharge, do not use the meter in a very dry environment, especially one in
which synthetic materials are present.