Display Shows Possible Cause(s) Possible Solution(s)
E-58 Temp. Error j The temperature is above or
below the proper range for
the meter.
j Move the meter to an area within the proper
temperature range indicated for test strip use in the
test strip package insert. Wait 5 minutes before
turning on the meter. Repeat the test. Do not
artificially heat or cool the meter.
E-59 Battery
Empty
j The batteries are extremely
low.
j Insert new batteries. See Chapter 7, “Care and
Maintenance.
E-60 Time/Date
Error
j The time and date settings
may be incorrect.
j You have changed the
batteries.
j Ensure the time and date are correct and adjust, if
necessary. See Chapter 5, “Changing Meter
Settings.
HI bG Warning j Your blood glucose may be
higher than the measuring
range of the system.
j If you are experiencing any of the common
symptoms of high blood glucose, contact your
healthcare professional immediately.
j Treat your high blood glucose as recommended by
your healthcare professional.
j Consider checking bG, ketones, and insulin.
j See the pump User Guide for additional solutions.
Incorrect Pump
Time/Date
j Pump time/date incorrect. j Set time/date on the pump (see pump User Guide).