Farbe: P 249 – Prüfmittelnummer: 126 X
Symbols, display messages and troubleshooting
149
Symbols, display messages and troubleshooting
10
This message/symbol
appears…
and means:
The bar code on the test strip drum
is damaged.
If you inserted the first drum from a
pack, remove the drum from the me
-
ter and return it to its container. In
-
sert the second drum from the same
pack in the meter and use all 17 test
strips. When it is time to replace the
drum, insert the first drum with the
damaged bar code. When the error
message is displayed again, press
the S button. The meter uses the bar
code from the preceding drum. The
drum and bottle symbols flash in the
display. The meter requests you to
perform a quality control check. You
may use all 17 test strips for testing.
The drum symbol continues to flash
in the display until you have inserted
a new drum.
This message/symbol
appears…
and means:
If you inserted the second or third
drum from a pack, press the S but
-
ton. The meter uses the bar code
from the preceding drum. The drum
and bottle symbols flash in the
display. The meter requests you to
perform a quality control check. You
may use all 17 test strips for testing.
The drum symbol continues to flash
in the display until you have inserted
a new drum.
If you inserted a test strip drum from
single-drum pack, you can no longer
use this drum. Insert a new drum.